Gas Adjustment Example
Enter Gas Type S.P. Function - Gas Type Adjust *C
Enter *Methane If CCC001, then Methane is only choice
Down Ethanol Press [E] if monitoring Ethanol
Mode Gas Type Return to S.P. Function - Gas Type Adjust
Down Range S.P. Function - Gas Range Adjust
Enter *100%LEL Select [E] to select or or to select another and
press [E]
4-20 mA Adjustment Example
Enter Calib S.P. Function - Calib Adjust *D
Enter Out: 4mA Use or keys to select 4 mA or 20mA
Enter 4mA Selects 4 mA
Enter ACK Momentary Acknowledge of new Set Point
Mode Calib S.P. Function - Calib Adjust
Down CalibOut S.P. Function - Output during Calibration Adjust
Enter Track Use or keys to select Track, Zero, C2.50mA, C4.00mA
Track = Output during calibration tracks the calibration gas,
C2.50mA = Output during calibration is 2.50 mA, C4.00mA =
Output during calibration is 4.0 mA
User defined
Enter *Track * = Current selection
RS-485 Adjustment Example
Enter Address S.P. Function - RS-485 Address Adjust *E
Enter Addr:016 Use or keys to enter new address
Enter ACK New address selected
Enter Address S.P. Function - RS-485 Address Adjust
Enter *38400 Press [E] to select or [] or [] to select another
Down Baud S.P. Function - RS-485 Baud Rate Adjust
Table 5 - 3 B: Set-Up Configuration