13.7 UPS System Replacement
If the UPS needs replaced, perform the following:
The part number for the UPS module is TE-W-201.
WARNING Before opening covers to perform maintenance, ensure unit is
powered down completely. To power down completely, remove mains power
and disconnect the battery at:
MAIN MENU
MAINTENANCE
MENU
BATTERY
DISCONNECT
1. Remove both the upper and lower left plates using a phillips
screwdriver.
2. The UPS module is located in the top left corner on the rear plate.
3. There are wires that enter on the top of the UPS and on the bottom of
the UPS.
4. On top there are (4) wires on TB2 terminals 1 through 4 that are the
alarm outputs for battery fail and battery discharge that route over to
the controller. Label each of these wires and unscrew the terminals
and remove the (4) top wires.
UPS Module
TE-W-201