13.13 Filter Temperature Probe Replacement
If the filter temperature probe needs replaced, perform the following:
The part number for the filter temperature probe is TE-W-152.
1. Unplug the filter temperature probe connector.
2. Unscrew the filter temperature probe from the thermowell on the filter
holder.
3. Install new filter temperature probe and connect to filter temperature
connector.
4. Acknowledge the filter temperature probe was unplugged alarm.
5. See section 7.2 to perform a calibration of the filter holder temperature
sensor.
Illustration showing filter temperature probe
13.14 Ambient Temperature Replacement
If the ambient temperature probe needs replaced, perform the following:
The part number for the ambient temperature probe with radiation shield is
TE-W-151.
1. Unplug the ambient temperature probe from the connector located on
the left side of the enclosure.