TE-WILBUR OPERATIONS MANUAL REV 002.15 06/11/2015
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 133 of 183
2. Remove the (2) 1/4-20x5/8” bolts that secure the radiation shield
bracket to the enclosure.
3. Install new radiation shield with ambient temperature probe onto the
enclosure with the (2) bolts.
4. Plug the temperature probe into the socket on the left side of the
enclosure.
5. Acknowledge the ambient temperature probe was unplugged alarm.
6. See section 7.1 to perform a calibration of the ambient temperature
sensor.
13.15 Solenoid Replacement
If the solenoid needs replaced, perform the following:
The part number for the solenoid is TE-W-212.
WARNING Before opening covers to perform maintenance, ensure unit is
powered down completely. To power down completely, remove mains power
and disconnect the battery at:
MAIN MENU
MAINTENANCE
MENU
BATTERY
DISCONNECT
1. Remove both the upper and lower left plates using a Phillips
screwdriver.
(2) Bolts