9. Connect the compression hose fittings to the inlet and outlet of the
solenoid.
10. Reinstall the upper and lower right plates and turn power on.
13.16 Pump Controller Replacement
If the pump controller needs replaced, perform the following:
The part number for the pump controller is TE-W-301with the TE-W-302
pump controller cable.
WARNING Before opening covers to perform maintenance, ensure unit is
powered down completely. To power down completely, remove mains power
and disconnect the battery at:
MAIN MENU
MAINTENANCE
MENU
BATTERY
DISCONNECT
1. Remove both the upper and lower left plates using a Phillips
screwdriver.
The pump controller is located on the back of the panel towards the bottom
left side.