TE-WILBUR OPERATIONS MANUAL REV 002.15 06/11/2015
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 77 of 183
8.5 System Information
The system information screen is where the software versions can be found
along with the total sampling runtime and the pump runtime. The pump
runtime can be reset here.
The system information screen can be found by following these keystrokes:
MAIN MENU
MAINTENANCE
MENU
SYSTEM
INFORMATION
System Information Screen
Total Sampling
Runtime
This is the total in hours that a sample has been running.
This value is not resettable.
Pump Runtime
This is the amount of time the pump has been running
either sampling, in manual mode or performing
calibrations. The pump runtime can be reset by pressing
the Reset Pump Runtime button. Pump rebuilds are
recommended at 3,000 hours and the operator is prompted
when the pump has 3,000 hours of runtime. See section
12.0 Maintenance for information on rebuilding the
pump.