TE-WILBUR OPERATIONS MANUAL REV 002.15 06/11/2015
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 94 of 183
10.3 Alarm Flowchart
The following shows how an alarm occurs, is acknowledged and returns to
normal operation with an example of the ambient temperature sensor being
unplugged.
Alarm
Occurrs
History
Log
Screen
Appears
Alarm
Condition
Resolved
Acknowledge
Button
Pressed
System
Back to
Normal
Ambient
Temperature is
unplugged
Operator is
notified. Alarm is
highlighted and in
RED
Operator plugs
Temperature
sensor back in
Operator Presses
button on History
log screen
Alarm turns black
indicating it is no
longer active
10.4 History Log Download
Data from the data log and the history log are saved from the internal memory
of the screen onto an SD located in the screen. The data that is saved to the
SD card can then be downloaded to the USB stick by inserting a USB stick
into the USB port on the front of the unit or by accessing the Data Download
screen by the following keystrokes:
MAIN MENU
MAINTENANCE
MENU
DATA DOWNLOAD
MENU
When inserting a USB drive into the USB port on the front the following
popup screen will appear:
USB Inserted popup Screen