TE-5170V VFC TSP 18 Operations Manual
Factor excel spreadsheet). It is highly recommended to download the
electronic excel spreadsheet and use spreadsheet features to complete
calculations. (If you do not have a G Factor then go to TE-5170V High Vol.
TSP excel spreadsheet on web site and then go to page 20)
G-Factor Excel Spreadsheet Instructions
The TE-5170V calibration worksheet can be found on our website at
www.tisch-env.com. If you have the G Factor that accompanies each
VFC, go to TE-5170V High Vol. TSP VFC with G-Factor, if ou do not
have a G Factor then go to TE-57V High Vol. T“P ecel
spreadsheet.
Note: Calibration orifices should be sent back to Tisch Environmental
for calibration on an annual basis per US EPA Compendium Method
IO-2.1 Part 7.3.2 Sampling of Ambient Air For Total Suspened
Particulate Matter (SPM) and PM10 Using High Volume (HV) Sampler
1. Enter the following information in the corresponding cells in the
worksheet:
Site Information
Location
The location of the instrument
Date
Current Date
Tech
Technician performing the calibration
Serial #
Serial number of the instrument, Pxxxx
VFC G
Factor
The g-factor of the VFC you are calibrating. This can be
found on the lookup table documentation (first page of this
doc) or the sticker located on the VFC.
Calibration Orifice Information
Make
The make of the orifice, typically Tisch Environmental
Model
The model number of the orifice, typically TE-5028A