39. correlation coeff. =
300511.74
61847.8
40. correlation coeff. =
6197744.8
61847.8
41. correlation coeff. = .9998
A calibration that has a correlation coefficient of less than .990 is not considered linear
and should be re-calibrated. As you can see from both worksheets we have 3 Qstd
numbers that are in the TSP range (1.1 - 1.7) and the correlation coefficient. is > .990 ,
thus a good calibration.