Loading...
TE-5170 TSP MFC 34 Operations Manual
5. Carefully slide quick disconnect tabs into tab slot of new brush.
6. Push brush carbon against armature until brush housing falls into brush slot on
motor.
7. Put brush holder clamps back onto brushes.
8. Make sure quick disconnect tabs are firmly seated into tab slot. Check field wires
for good connections.
9. Assemble motor after brush replacement by placing housing over and down on the
motor (at same time pull power cord out of housing), being careful not to pinch any
motor wires beneath the motor spacer ring.
10. Secure power cord with the cord retainer cap.
11. Replace blower motor flange on top of motor making sure to center gasket.
Assemble together with filter holder. Lower filter holder and blower motor down
through top support pan on shelter.
Loading...
Terms of Use | Privacy Policy | DMCA Policy