Loading...
TE-1000 PUF 23 Operations Manual
36. correlation coeff. =
)] - [(2)] . - [(.
5-
)(
402.2466056.513020882306313
77549.8
1379.43
37. correlation coeff. =
)]2 - )][(2. - [(.
) -(
402.466056.513020882306313
62758.877549.8
38. correlation coeff. =
)2036.5)(0042248(.
791.14
39. correlation coeff. =
0219841..
14791.
40. correlation coeff. =
82703.14
791.14
41. correlation coeff. = .9975699
A calibration that has a correlation coefficient of less than .990 is not considered
linear and should be re-calibrated. Since the correlation coeff. is > .990 , we have a
good calibration.