TE-1000 PUF 30 Operations Manual
Maintenance and Care
A regular maintenance schedule will allow a monitoring network to operate for
longer periods of time without system failure. Our customers may find that the
adjustments in routine maintenance frequencies are necessary due to the
operational demands on their sampler(s). We recommend that the following
cleaning and maintenance activities be observed until a stable operating history
of the sampler has been established.
TE-1000 PUF
1. Make sure all gaskets (including motor cushion) are in good shape and that
they seal properly.
2. The power cords should be checked for good connections and for cracks
(replace if necessary).
CAUTION: DO NOT allow power cord or outlets to be immersed in water!
3. Inspect the filter screen and remove any foreign deposits.
4. Inspect the filter holder frame Teflon gasket each sample period and make
sure of an airtight seal.
5. Check or replace TE-33384 or TE-33378 motor brushes every 400 to 500
running hours.
6. After replacing brushes two times, a new motor (TE-116336 or TE-116125)
must be used.
7. Make sure elapsed time indicator is working properly.