TE-5170D-BL TSP MFC 25 Operations Manual
37. correlation coeff. =
)] - )][( - [(
) -(
6928.8928977.915965216.10973412.10
39888.30801735.317
38. correlation coeff. =
))((2842.231321252.0
61847.8
39. correlation coeff. =
300511.74
61847.8
40. correlation coeff. =
6197744.8
61847.8
41. correlation coeff. = .9998
A calibration that has a correlation coefficient of less than .990 is not considered linear and
should be re-calibrated. As you can see from both worksheets we have 3 Qstd numbers that are
in the TSP range (1.1 - 1.7) and the correlation coefficient. is > .990 , thus a good calibration.