11. Replace blower motor flange on top of motor making sure to center
gasket.
**IMPORTANT** To enhance motor life:
Change brushes before brush shunt touches armature.
Seat new brushes by applying 50% voltage for 10 to 15 minutes, the
TE-5075 brush break in device allows for the 50% voltage.
TE-116311
110v Motor for TE-5200
TE-33384(green)
110v Motor Brush
TE-116312
220v Motor for TE-5200
TE-33378(brown)
220v Motor Brush