Loading...
TE-5200 9 Operations Manual
Calibration Kits
The TE-5028 is the preferred method to calibrate the TE-5200 TSP Tripod
High Volume Air Sampler. It simulates change in the resistance by merely
rotating the knob on the top of the calibrator. The infinite resolution lets the
technician select the desired flow resistance. The TE-5028 calibration kit
includes: carrying case,  slack tube water manometer, adapter plate, ’
piece of tubing, and TE-5028A orifice with flow calibration certificate.
Each TE-5028A is individually calibrated on a primary standard positive
displacement device (Rootsmeter) which is directly traceable to NIST.
** It is recommended that each calibrator should be re-calibrated
annually for accuracy and reliability.
Parts
1. Tri-Pod Shelter Box - 27" x 21" x 20" 36 lbs
Tri-Pod Pan
TE-5200 110volt, 60hz
TE-5200X 220volt, 50hz
TE-5200XZ 220volt, 60hz