TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 100 of 185
2. Close the filter holder
3. Remove the PM-10 inlet and install the TE-L30 flow rate adapter on
the top of the downtube.
4. Close the valve on the flow rate adapter to plug the air flow.
5. Press the Select External Leak Check button on the Leak check
selection screen.
6. At the Perform External Leak Check Next Steps screen, the screen will
show the TE-L30 adapter closed and the filter holder closed. Press
next step and then press the Final Step button.
7. You will now be at the Leak Check Control Screen.
Leak Check Control screen
Picture of the TE-L30
Downtube adapter placed
on top of the downtube with
the valve in the closed
position