15.0 Communications
This section outlines the communication options, configuration and operation
available with the TE-Wilbur sampler.
15.1 Introduction
The TE-Wilbur has a built-in RJ-45 Ethernet connection that allows for remote
access of the system, a web browser interface and the ability to download data
remotely. Also, the TE-Wilbur sampler has a Modbus slave port that allows
connection to a telemetry system for remote data logging and viewing and
connection to the Tisch Environmental FRM-CAL Calibrator.
15.2 Ethernet Connection
The TE-Wilbur sampler has a standard RJ-45 Ethernet connection that is
located in the weatherproof box on the side. This connection is routed through
to the touchscreen where the connection is made.
If connection of the sampler is made to a switch, then a standard RJ45 CAT5/6
patch cable can be used.
If connecting this port directly to a PC an RJ45 CAT5/6 crossover cable must
be used.
The system is set for a default IP address as follows:
IP Address 192.168.1.123
Subnet Mask 255.255.255.0
No gateway is set
In order to connect to the sampler via the Ethernet port your PC must also be
set to the same subnet, for example:
IP Address 192.168.1.124 (notice the IPs cannot be the same)
Subnet Mask 255.255.255.0
In this example the subnet is ‘192.168.1’ and any addresses from 1 to 255 can
be used except the ‘.123’ that is being used by the sampler.