6. Using a soft brush, cloth and cotton swabs, lightly scrub all interior
surfaces and the bug screen with distilled water and / or general purpose
cleaner.
7. Check the o-ring around the upper unit that seals the upper and lower
pieces for wear and damage. Replace if worn or damaged
8. Check the 2 o-rings in the bottom of the inlet for wear and damage.
Replace if worn or damaged.
9. Place a small amount of o-ring grease around the o-ring that seals the
upper and lower units.
10. Screw the upper and lower unit back together carefully.
11. Place a small amount of o-ring grease around the 2 o-rings on the
bottom.
12. Screw the jar back into the inlet and place the inlet back onto the
downtube.
The PM10 Inlet has the following O-rings that should be inspected for wear and
replaced as needed.
Quantity
Description
Part Number
2
Exit Adapter O-Rings
TE-W-022
1
Impactor Nozzle O-Ring
TE-W-023
2
Downtube O-Rings
TE-W-021
These can also be ordered together in a kit by part number TE-W-020.
There are also O-rings located in the black downtube adapter and those can be
ordered using part number TE-W-024.
5.3 TE-PM2.5C Cyclone
The TE-PM2.5C is a PM2.5 particle fractionator that utilizes a cyclonic effect
to separate ambient particles that have an aerodynamic diameter less than
2.5um. The cyclone is placed directly underneath the PM10 inlet to achieve
PM2.5 collection.