Loading...
TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 60 of 185
To calibrate the filter temperature sensor perform the following:
1. Obtain a calibrated temperature device such as the Tisch FRM-CAL low
volume calibrator.
2. Allow the calibrated temperature device to reach equilibrium with the
ambient air.
3. Place the temperature device into the filter holder holding it at the tip of
the filter temperature RTD.
4. The temperature calibration can be performed by following these
keystrokes:
MAIN MENU
CALIBRATION
FILTER TEMPERATURE
CALIBRATION
5. Press the box and enter the temperature reading from the calibrator.
6. Press UPDATE to update the temperature.
Filter Temperature RTD
Loading...
Terms of Use | Privacy Policy | DMCA Policy
2006-2020 Rsmanuals.com