TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 70 of 185
7.9 TE-Wilbur Calibration Worksheet Instructions
1. Fill out the following information in the first box:
Date / Time
Current date and time
Serial Number
Serial Number of the TE-Wilbur unit
Technician
Technician performing the calibration
Calibrator Make/Model
The make and model of the calibrator you are using to perform
the calibration
S/N:
The serial number of the calibrator you are using to perform the
calibration
Calibrator Due Date
The date the calibrator is due for re-certification, this is typically
1-year after the last certification
2. The second box is the ambient temperature, filter temperature and barometric pressure calibration.
Start an ambient temperature calibration. Refer to Section 7.1 of the Operations Manual for
performing ambient temperature calibration. Enter the following information:
As Found
The current reading of ambient temperature of the TE-Wilbur
system
Calibrator Reading
The current reading of the calibrator ambient temperature
As Left
3. Perform a filter temperature calibration. Start a filter temperature calibration. Refer to Section 7.2
of the Operations Manual for performing an filter temperature calibration. Enter the following
information:
As Found
The current reading of filter temperature of the TE-Wilbur
system
Calibrator Reading
The current reading of the calibrator filter temperature
As Left
The value of filter temperature the TE-Wilbur system was left at.
This should always be the calibrator reading