Loading...
TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 84 of 185
9.0 Data Logging / Operational Data
This section explains the data functions with TE-Wilbur, how data is collected,
stored in the datalog and displayed on the screen.
9.1 Operational Data
The operational data can be found by following these keystrokes:
MAIN MENU
OPERATIONAL
DATA
There are (3) screens of operational data. On these screens the following
information is presented to the operator:
Volumetric Air
Flow (Lpm)
The volumetric air flow (Qactual) of the system in real
time and the maximum and minimum volumetric flow
rate updated every 30 seconds.
Barometric
Pressure (mmHg)
The barometric pressure in millimeters of mercury in
real time and the maximum and minimum barometric
pressure updated every 30 seconds.
System Pressure
(“H2O)
The system vacuum pressure in inches of water in real
time and the maximum and minimum system pressure
updated every 30 seconds.
Filter Temp
(Deg C)
The filter temperature in degrees Celsius in real time and
the maximum and minimum filter temperature updated
every 30 seconds.
Ambient Temp
(Deg C)
The ambient temperature in degrees Celsius in real time
and the maximum and minimum ambient temperature
updated every 30 seconds.
Temperature
Variance
(Deg C)
The difference between the ambient temperature and the
filter temperature in real time and the maximum and
minimum temperature variance updated every 30
seconds.
Total Airflow
(Liters / m3)
The total air flow shown in cubic meters and in liters.