TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 9 of 185
Although this system is not intended for running a full 24 hour sample without
mains power, the system includes an external power connection that can be used
to connect larger external batteries, solar, wind or any other 24VDC alternative
energy sources. Help menus are included on the screen to assist the operator
with calibration, leak checking and other features of the system. The system
provides remote Ethernet control and monitoring from any standard browser
and a Modbus interface for remote telemetry connections.
1.2 US EPA FRM Designation
The US EPA has designated the Tisch Environmental TE-Wilbur10 and TE-
Wilbur2.5 as a Federal Reference Method (FRM) and a Federal Equivalency
Method (FEM) per the table below for the measurement of particulate matter in
ambient air based on the requirements of 40 CFR Part 53.
This unit is a gravimetric method and meets all requirements of 40 CFR Part 50
Appendix L and has been tested and meets the requirements of 40 CFR Part 53.
Method
Configuration
Designation
Designation Number
PM10
TE-PM10-D PM10 head
FRM
RFPS-0714-216
PM2.5
TE-PM10-D PM10 head
with VSCC Cyclone, WINS
Impactor or TE-PM2.5C Cyclone
FRM
RFPS-1014-219
PM2.5
TE-PM10-D PM10 head
TE-PM2.5C Cyclone
FEM Class II
EQPS-0415-223
PM10-2.5
*Requires
collocation of (2)
samplers
Sampler 1: (PM2.5)
TE-PM10-D PM10 head
with VSCC Cyclone, WINS
Impactor or TE-PM2.5C Cyclone
Sampler 2: (PM10)
TE-PM10-D PM10 head
FRM
RFPS-1014-220
PM10-2.5
*Requires
collocation of (2)
samplers
Sampler 1: (PM2.5)
TE-PM10-D PM10 head
with TE-PM2.5C
Sampler 2: (PM10)
TE-PM10-D PM10 head
FEM Class II
EQPS-0415-224
Loading...
Terms of Use | Privacy Policy | DMCA Policy
2006-2020 Rsmanuals.com