Loading...
TE-WILBUR OPERATIONS MANUAL REV 002.22 4/27/2017
Tisch Environmental 1-TSP-AND-PM10 sales@tisch-env.com www.tisch-env.com Page 94 of 185
1 in 6 sample selected
1in6 sample has been selected and engaged.
1 in 12 sample selected
1in12 sample has been selected and engaged.
Custom sample selected
A custom sample has been selected and engaged.
Flow variance alarm during
sample
The flow varied by more than 10% for 10 minutes during a sample
this will shut down a sample.
5% Flow variance warning
The flow varied by more than 5% for 5 minutes, will give a
warning for the sample.
Sample less than 23or
greater than 25 hours
The sample time was less than 23 hours or greater than 25 hours,
most probably due to a complete loss of power. Will give a
warning for a sample.
Ambient temperature
unplugged or failed
The ambient temperature sensor was unplugged, dropped below
-90 Deg C, went above +60 Deg C or there was an internal problem
with the controller this will shut down a sample.
Filter temperature
unplugged or failed
The filter temperature sensor was unplugged, dropped below
-90 Deg C, went above +60 Deg C or there was an internal problem
with the controller this will shut down a sample.
Flowmeter unplugged or
failed
The flowmeter was unplugged, went above 25 Lpm, below 0 Lpm
or there was an internal problem with the controller this will shut
down a sample.
Barometric press sensor
unplugged or failed
The barometric pressure sensor was unplugged, went below 450
mmHg, went above 850mmHg or there was an internal problem
with the controller this will shut down a sample.
System pressure sensor
unplugged or failed
The pressure sensor went above 15 psi of vacuum, fell below 0 psi
or there was an internal problem with the controller.
Sample shutdown due to
alarm condition
The current sample was terminated due to an alarm.
5 Deg temp variance for 30
minutes
The difference between the ambient temperature and the filter
temperature was greater than 5 Degrees Celsius
Barometric pressure
updated
The barometric pressure sensor was calibrated and updated to a
new value