TE-CD-TSP 24 Operations Manual
Note: Calibration orifices should be sent back to Tisch Environmental for calibration on
an annual basis per US EPA Compendium Method IO-2.1 Part 7.3.2 Sampling of Ambient
Air For Total Suspened Particulate Matter (SPM) and PM10 Using High Volume (HV)
Sampler
1. Enter the following information in the corresponding cells in the worksheet:
Site Information
Location
The location of the instrument
Date
Current Date
Tech
Technician performing the calibration
Serial #
Serial number of the instrument, Pxxxx
VFC G
Factor
The g-factor of the VFC you are calibrating. This can be found on
the lookup table documentation (first page of this doc) or the
sticker located on the VFC.
Calibration Orifice Information
Make
The make of the orifice, typically Tisch Environmental
Model
The model number of the orifice, typically TE-5028A
Serial #
The Serial number of the calibration orifice you are using
Qa Slope
(m)
The Qa slope of the calibration orifice you are using. This is
found on the calibration documentation provided with the
calibration orifice
Qa Int (b)
The Qa intercept of the calibration orifice you are using. This is
found on the calibration documentation provided with the
calibration orifice
Calibration
Due Date
The date that the calibration of the orifice is due. Orifices
should be calibrated on an annual basis. Call Tisch
Environmental at 1-TSP-AND-PM10 to schedule a calibration.
Ambient Conditions
Temp (Deg
F)
Enter the current ambient temperature at calibration, Ta in
Degrees K and Ta in degrees C will be calculated by the
spreadsheet