TE-CD-TSP 48 Operations Manual
DATE Menu Item
The DATE menu item is used to specify the start date of the timer. The date must be entered in the
MMDDYY format. Leading zeros are required (For example, to sample on January 1, 2007, enter
010107).
TIME Menu Item
The TIME menu item is used to specify the start time of the timer. The time must be entered in the
HHMM format. Leading zeros are required (For example, to sample at 09:30 AM, enter 0930).
DURATION Menu Item
The DURATION menu item is used to specify the length of time the unit should sample. The duration
must be entered in the HHMM format. Leading zeros are required (For example, to sample for 1 hour,
enter 0100). This is the intended length of the sample. Power failures or other interruptions will not
extend the sample time. For example, if the timer is set to turn on at 12:00 and run for 24 hours, but the
power fails for 2 hours. The timer will stop at 12:00 the following day.
REPEAT Menu Item
The REPEAT menu item is used to specify the interval of time between sample start times. If set to zero
(NONE), the repeat feature is disabled and the timer will only execute once. Several standard repeat
intervals are selectable. For example, the “1 IN 3” selection will set the repeat interval for sampling
every third day (a new sample would start every 72 hours). The CUSTOM selection allows for non-
standard repeat intervals to be entered. When entering non-standard repeat intervals, the HHMM format
must be used. Leading zeros are required (For example, to sample for 1 hour, enter 0100).
SAVE and EXIT Menu Item
The SAVE and EXIT menu item saves the current setup and returns to the Main Status Display.
STOP and EXIT Menu Item