TE-CD-TSP 52 Operations Manual
FLAGS
COMPLETED: Y
EXECUTING: N
ABORTED: N
EXPIRED: N
FLOW RANGE: N
POWERFAIL: N
Qstd: N
The third screen (shown above) displays flags associated with the timer event. The flags are described
below.
FLAG
Description
COMPLETED
Indicates that the Timer Event has been completed
EXECUTING
Indicates that the Timer Event is currently executing
ABORTED
Indicates that the Timer Event was aborted by the user
EXPIRED
Indicates that the Timer Event expired before it could be started.
This can happen if the unit is without power during the entire
timer event.
FLOW RANGE
Indicates that the sampling flow rates exceeded +/- 10 percent
of the set flow rate for a period exceeding 2 minutes. This
indicates that the system was not able to control the flow at the
desired set point and the sample event was stopped for this
reason. Examine the system for disconnected tubing, clogged
orifices, proper sensor operation, calibration errors, and filter
media requirements to determine the source of the problem.
POWERFAIL
Indicates that a power failure occurred during the Timer Event.
Qstd
Indicated that the Timer Event is using Qstd for flow control.