TE-CD-TSP 59 Operations Manual
CONFIGURATION
The CONFIGURE menu (shown below) allows for various operating parameter of the unit to be
configured. Details are provided below.
CONFIGURE
SET DATE: 04-20-07
SET TIME: 12:11:18
FLOW CONDITIONS: AMB
FLOW RATE: 40
LCD CONTRAST: 32
LOG INTERVAL: 5
TIMER MODE: TIMER
Setting the Date and Time
The SET DATE and SET TIME menu items allow the battery backup real-time clock to be
programmed. This clock provides the time base used for operation with the timer. The clock
automatically adjusts for leap years, but does not automatically adjust for daylight savings time. If the
date needs to be set, it should be set first. The date should be entered in the format MMDDYY with
leading zeros. For example, January 3rd, 2007 would be entered as 010307. The time should be entered
in the 24-hour format HHMM with leading zero. For example, 1:14 AM would be entered as 0114,
while 1:14 PM would be entered as 13:14.
Setting the Flow Conditions
The FLOW CONDITIONS menu item selects whether the unit will operate in ambient (AMB)
conditions or standard (STD) conditions flow. Standard conditions are defined to be 760 mmHg and
25°C
Setting the Flow Rate
The FLOW RATE menu item determines the operating flow rate for the system and is measured in
CFM. For VFC+ systems changing the flow rate will not change the flow rate, the flow rate is
determined by the VFC device not the VFC+ controller. This setting is only indented to be displayed on
the data.csv file for reference only. CHANGING THIS WILL NOT CHANGE THE SYSTEMS
FLOWRATE.