Pamb
The Pamb menu item displays the ambient pressure (uncorrected barometric pressure) being measured
by the system. The value in parentheses is the raw analog to digital converter (ADC) value (range of 0 -
4095), while the number to the right represents the calibrated value. An ADC value of 0 or 4095
indicates that the sensor is saturated. If this is observed, check all system connections. If the problem
persists, contact technical support for further instruction.
Pdif
The Pdif menu item displays the pressure drop (across the filter) being measured by the system. This
sensor is not used in the flow control calculation on the VFC +. The value in parentheses is the raw
analog to digital converter (ADC) value (range of 0 - 4095), while the number to the right represents the
calibrated value. An ADC value of 0 or 4095 indicates that the sensor is saturated. If this is observed,
check all system connections looking for kinks or cuts in the tubing. If the problem persists, contact
technical support for further instruction.
Qstd (or Qamb)
The Qstd (or Qamb) menu item displays the flow rate being measured by the system. The menu item
label changes between Qstd and Qamb to indicate the current flow conditions as determined by the MTR
CONTROL setting. This is where the Gfactor on the VFC device will be entered.
Motor
Use this function to manually toggle the motor on/off.