Maintenance and Care
A regular maintenance schedule will allow a monitoring network to operate
for longer periods of time without system failure. Our customers may find
that the adjustments in routine maintenance frequencies are necessary due
to the operational demands on their sampler(s). We recommend that the
following cleaning and maintenance activities be observed until a stable
operating history of the sampler has been established.
TE-PNY1123
1. Make sure all gaskets (including motor cushion) are in good shape and
that they seal properly.
2. The power cords should be checked for good connections and for cracks
(replace if necessary).
CAUTION: DO NOT allow power cord or outlets to be immersed in water!
3. Inspect the filter screen and remove any foreign deposits.
4. Inspect the filter holder frame gasket each sample period and make sure
of an airtight seal.
5. Check or replace TE-33384 motor brushes every 400 to 500 running
hours.
6. After replacing brushes two times, a new motor (TE-116336 or
TE116125) must be used.
7. Make sure elapsed time indicator is working properly.