TE-5170DV TSP VFC 19 Operations Manual
Air For Total Suspened Particulate Matter (SPM) and PM10 Using High Volume (HV)
Sampler
1. Enter the following information in the corresponding cells in the worksheet:
Site Information
Location
The location of the instrument
Date
Current Date
Tech
Technician performing the calibration
Serial #
Serial number of the instrument, Pxxxx
VFC G Factor
The g-factor of the VFC you are calibrating. This can be found on the lookup table
documentation (first page of this doc) or the sticker located on the VFC.
Calibration Orifice Information
Make
The make of the orifice, typically Tisch Environmental
Model
The model number of the orifice, typically TE-5028A
Serial #
The Serial number of the calibration orifice you are using
Qa Slope (m)
The Qa slope of the calibration orifice you are using. This is found on the calibration
documentation provided with the calibration orifice
Qa Int (b)
The Qa intercept of the calibration orifice you are using. This is found on the calibration
documentation provided with the calibration orifice
Calibration Due Date
The date that the calibration of the orifice is due. Orifices should be calibrated on an
annual basis. Call Tisch Environmental at 1-TSP-AND-PM10 to schedule a calibration.
Ambient Conditions
Temp (Deg F)
Enter the current ambient temperature at calibration, Ta in Degrees K and Ta in
degrees C will be calculated by the spreadsheet
Barometric Pressure
2. Enter the calibration information by performing each calibration point and
entering the following information into each corresponding row for each
point: