TE-5170DV TSP VFC 9 Operations Manual
Calibration Kits
The TE-5028 is the calibration kit available for use with the TE-5170DV TSP VFC High
Volume Air Sampler.
The TE-5028 is the preferred method to calibrate the TE-5170DV VFC TSP High
Volume Air Sampler. It simulates change in the resistance by merely rotating the knob
on the top of the calibrator. The infinite resolution lets the technician select the desired
flow resistance. The TE-5028 calibration kit includes: carrying case, 3 slack tube
water manometer, adapter plate, 3’ piece of tubing, and TE-5028A orifice with flow
calibration certificate. Optional electronic manometer is available.
Each TE-5028A is individually calibrated on a primary standard positive displacement
device (Rootsmeter) which is directly traceable to NIST.
** It is recommended by the EPA that each calibrator should be re-calibrated
annually. (1998 Code of Federal Regulations Parts 50 to 51, Appendix B to Part
50, Reference Method for the Determination of Suspended Particulate Matter
in the Atmosphere, 9.2.5 page 29.)