TE-5000-BL TSP 19 Operations Manual
m =
 
( x)( y)
xy - n
x ; b = y - mx
x - n
2
2
 
where:
n = number of observations
y = y/n; x = x/n
= sum of
The equation for the coefficient of correlation (r) is as follows:
r =
 
( x)( y)
xy - n
x-
x
n y-
y
n
22
 
2 2
where:
n = number of observations
= sum of
The acceptable operating flow range of a TSP sampler is 1.1 to 1.7 m3/min (39 to 60
CFM). Looking at the worksheet column Qstd, the flow rates that are within this
range can be identified along with the chart reading that represents them. For
instance if you wanted to set this sampler at 1.451 m3/min (51.24 CFM) (Make sure
the variac (TE-5010) is plugged in and a filter is in place) you would turn the shaft
lock on the variac (TE-5010) until the continuous flow recorder read 57 on the
chart. By making sure that the sampler is operating at a chart reading that is within
the acceptable range, it can be assumed that valid TSP data is being collected.