89
SETTING FUNCTIONS ON SETUP SCREEN
KRT
"KRT" contains settings related to output from the internal printer.
TONO/PACHO
"TONO/PACHO" contains settings related to output from the internal printer.
Descriptions Options Details Initial value
KRT print order
D/mm KRT data is printed as follows,
D (corneal refractive power)/mm (corneal curvature). D/mm
mm/D KRT data is printed as follows,
mm (corneal curvature)/D (corneal refractive power).
KRT format ALL All the measurement value is printed. ALL
AVG Printout only typical value.
KRT style
HV Display style of KRT measurement results is set to
HV (horizontal/vertical). R1R2
R1R2 Display style of KRT measurement results is set to
R1R2 (flat/steep meridian).
KRT print format HV KRT measurement result is printed in simple format. R1R2
R1R2 KRT measurement result is printed in full format.
KRT average OFF Do not print KRT average value. ON
ON Print KRT average value.
KRT cylinder OFF Do not print kerato-cylinder value and axial angle. ON
ON Print kerato-cylinder value and axial angle.
Cornea diameter OFF Do not print corneal diameter. ON
ON Print corneal diameter.
Descriptions Options Details Initial value
Printer order
R/L The order is right eye and left eye regardless of the TONO
measurement value and PACHO measurement value.
SIMPLE
DATA TONO measurement value and PACHO
measurement value are separately printed.
SIMPLE Print in the SIMPLE format.
Measure correction OFF Corrected measurement value is not printed. ON
ON Corrected measurement value is printed.
mmHg Display on hPa OFF mmHg is not printed on hPa. ON
ON mmHg is printed on hPa.
IOP ADJ Formula
OFF Center CCT Base and Adjustment Coefficient for
IOP ADJ formula are not printed. ON
ON Center CCT Base and Adjustment Coefficient for
IOP ADJ formula are printed.
Loading...