PREPARATIONS BEFORE USE
PREPARATIONS BEFORE USE
Before using this instrument, carry out the daily check. (Refer to "DAILY CHECK" on P.103.)
STERILIZED CAP: Common
Sterilize caps as follows:
Method: High pressure steam sterilization
Condition: Leave cap in the saturated steam (134±4°C) for 20 minutes.
PREPARATIONS FOR STERILIZING THE FRONT LENS UNIT: OFFISS, OFFISS Lite
Place the front lens unit, peripheral prism and anterior eye section observation lens in the sterilized tray
and sterilize them as follows:
Method: EOG (ethylene oxide gas) sterilization
Condition: Make sure the devices used are properly sterilized.
TURNING ON THE POWER: Common
1Connect the power plug to a wall-outlet.
2Switch the power switch " " (ON).
On the light intensity display window, the version information of the software installed in this
instrument is displayed for several seconds. (Display example: 1.03) If the software of
Version 1.02 or previous is installed, the version information will not be displayed.
3Make sure that the power indicator lights up green.
4Make sure that no error code is displayed on the light intensity display window. (Refer to
"ERROR CODES" on P.88.)
WARNING Make sure no-one is too close to the instrument before moving the arm.
Anyone touching the instrument may be injured.
NOTE Make sure the electricity is properly grounded.