Communication format Mode 2:
Model name, Type No. 10 byte
Time/Date 19 byte
Measurement value (right or left eye) average data 9 byte
If ERRs only, transmission is not done.
(Example)
Average Value Displayed in Decimal Format
Average Value Displayed in Integer Format
For the Decimal Format, the “measurement value” points are similar to those in “Com-
munication Format Mode 1, 3”.
Communication format Mode 4, STD 1:
Data Transmission
Model name, Type No. 15 byte
Machine No. 2 byte
ROM version 10 byte
ID No. 13 byte
Work ID No. 13 byte
Machine work ID No. 4 byte
Time/Date 20 byte
R (right eye) average data 9 byte
L (left eye) average data 9 byte
(Example)
Average Value Displayed in Decimal Format
Average Value Displayed in Integer Format
For the Decimal Format, the “measurement value” points are similar to those in “Com-
munication Format Mode 1, 3”.