The function of each switch is as follows:
Switch
Description
[AUTO/MANU.]
Auto/Manual
Switch for selection between [AUTO] and [MANU].
[AUTO]: Optimum measurement range is automatically selected
according to illuminance level.
[MANU.]: Measurement range is manually set.
Display range 2.1 Auto-range measurement and display range
[MANU. RANGE]
Manual Range
Changes the measurement range to the next manually.
Measurement range are switched in turn as follows;
1→2→3→456
Display range 2.2 Manual-range measurement and display range
[C.C.F.]
Color Correction Factor
Sets and confirms the value of the correction factor in correction factor
mode. The IM-600 can memorize one correction factor data.
[INV.]
Inverse
Changes the assignment of function to keys.
([cd] to [CALL], [0 9] to [ ], [SHIFT] to [CAL.], [SET] to
[START/STOP])
[]
Switches to percent mode in which the instrument measure the ratio of
the difference between measured illuminance data and reference
illuminance data. (% mode)
In addition, sets reference illuminance data.
[]
Switches to illuminance difference measurement mode in which the
instrument measure the difference between measured illuminance data
and reference illuminance data. ( mode)
In addition, sets reference illuminance data.
[lx]
Enters illuminance measurement mode. (lx mode)
[lxh]
Enters integral illuminance measurement mode.(lxh mode)
[cd]
Enters luminous intensity mode (cd mode).
[0 - 9]
Increments the numerical value when inputting reference data and
correction factor.
[SHIFT]
Moves the cursor down digit when inputting reference data and
correction factor.
[SET]
Confirms your entry by pressing this button in each setting.
[CALL]
([INV.]+[cd])
In the % and the mode, shows entered reference illuminance data
for 3 seconds.
In the cd mode, shows entered distance data for 3 seconds.
[ ]
([INV.]+[0 - 9])
Sets decimal point when entering the reference illuminance data,
distanced data, and correction factor in each mode.
[CAL.]
Calibration
([INV.]+[SHIFT])
Starts zero-adjustment manually. Zero-adjustment provides constant
sensitivity of photo detector and steady measurement. Each time
turning on the power, zero-adjustment starts automatically without
pushing the [CAL.] switch.
[START/STOP]
([INV.]+[SET])
Starts or stops integral illuminance measurement in the lxh mode.