22
Display range and resolution in auto-range unit: lx
Range
Effective measuring range
Resolution
MIN
MAX
Range 1
0.005
9.990
0.005
Range 2
7.50
99.90
0.05
Range 3
75.0
999.0
0.5
Range 4
750
9990
5
Range 5
7500
99900
50
Range 6
75000
999000
500
4
Following remark appear at the moment range change to another step.
Light should be uniformly distributed over whole surface of the detector window,
otherwise measured data may not be correct.
The sensitivity is not uniform over the whole area of the detector.
If the readout exceeds 999,000lx, a code E2 appears.
By switching the response switch, you can select the detector responding
speed. Select the responding speed [FAST] or [SLOW] depending on the light
source to be measured.
Response 1.2 Names and functions of parts
If the readout is smaller than 0.005lx, the figure 0 appears.
When the unit is "fc", the measurement range and resolution are obtained by
multiplying the above values by 0.0929.
Memo