A measured value appears in the display, read the value when it becomes stable.
To fix the readout, set the hold switch to ON.
5
For continuous measurement, set the hold switch to OFF, and restart measurement.
Display range and resolution in manual-range unit: lx
Range
Effective measuring range
Resolution
MIN
MAX
Range 1
0.005
9.990
0.005
Range 2
0.05
99.90
0.05
9.990
9.990
99.90
Range 3
0.5
999.0
0.5
9.990
9.990
99.90
99.90
999.0
Range 4
5
9990
5
9.990
9.990
99.90
99.90
999.0
999.0
9990
Range 5
50
99900
50
99.90
99.90
999.0
999.0
9990
9990
99900
: 50
Range 6
500
999000
500
999.0
999.0
9990
9990
99900
99900
999000
If the readout is exceeds the display range, a code E2 appears.
By switching the response switch, you can select the detector responding
speed. Select the responding speed [FAST] or [SLOW] depending on the
measuring light source to be measured.
Response 1.2 Names and functions of parts
If the readout is smaller than 0.005lx, the figure 0 appears.
When the unit is "fc", the measurement range and resolution are obtained by
multiplying the above values by 0.0929.