28
Enter the correction factor in the range 0.001 to 1000.
To cancel the correction factor, enter 1.000 as the correction factor.
After completion of the C.C.F. setting mode, the system returns to the previous
measuring mode.
Once correction factor is activated, Readout is multiplied by the correction
factor.
Readout = raw measured value x Correction factor (C.C.F.).
If the readout exceeds 999900 due to entering too large a correction factor, the
code E4 appears.
During measurement of the integral illuminance, the correction factor cannot be
changed.
Correction factor data is saved after turning off.
3.4.1 SETTING PROCEDURE (Ex:To enter 1.2)
1
The currently entered correction factor appears for 3 seconds after pushing [C.C.F.]
switch.
2
Push the [C.C.F.] switch again within 3 second after the above process, then the
screen turns to the screen for numeric value entry.
3
Push the [0 - 9] switch to change the blinking digit to 1.
4
Push the [INV.] switch, and check if the mark INV appears.
Memo