30
3.5 DIFFERENCE MEASUREMENT (
MODE)
Press the [
] switch to enter the

mode. In this mode, the difference between the
reference value and measured value appears.
(An example showing that the difference is +1880lx)
Difference = (C.C.F. x Illuminance value) Reference value
(C.C.F. x Illuminance value): Readout in the lx(fc) mode
If a reference value is not entered, the screen changes to the numeric entry
screen for reference value.
How to enter a reference value (based on the measured value)
1
In the lx mode, make measurement of illuminance for the reference illuminance data.
2
When measured value becomes stable, push the hold switch to ON, and fix the
display.
3
Push the [SET] switch. The measured value is set as the reference value.
4
Push the hold switch to OFF.
5
Push the [
] switch to shift to
mode.
How to enter the reference value (setting a given value as reference)
1
Push the [
] switch to shift to
mode.
2
Enter a numerical value and push the [SET] switch to activate the reference value.
Reference value might be entered from 0.001 to 999,900.
If the value 0 is entered as the reference value, a code E7 appears.
Once the reference value is entered, it continues to be active until it is changed
or the power is turned off.
Entering numerical value 3.1 NUMERIC ENTRY
Setting procedure 3.4.1 SETTING PROCEDURE
Memo
Memo