32
5
Push the [SHIFT] switch again to move the blinking digit to a lower position.
6
Push the [INV.] switch, and check if the mark INV appears.
7
Push the [ ] ([0 - 9])switch to set the decimal point.
8
Push the [SET] switch to complete the setting.
To amend the entered value before completing the setting, push the [SHIFT]
switch a few times to specify the digit to be amended. The only figure able to
be entered in the 5th and 6th digit is 0 (zero).
When the reference value is set with the [SET] switch, the reference value
appears for about 3 seconds and then the system returns to the previous mode.
3.5.2 HOW TO CHECK THE SETTING
Push [INV.] switch and then push [CALL] switch to confirm the entered reference
illuminance data and distance data in each mode.
The reference values for the
mode are the same as for the % mode.
[CALL] is valid only during operating in
 and cd mode.