4. COMMUNICATION WITH PC
4.1 COMMUNICATION COMMAND
This instrument can communicate with PCs. This chapter describes the commands for creating
communication program with the IM-600/IM-600M on user own.
The following are the communication commands:
Command name
Function
STR0
Sets measurement range to the auto, and obtains measured data.
STRn
Sets measurement range to the manual and designates a range, and then
obtains measured data
n: 1 6
ISR0
Sets measurement range to the auto, and starts integral illuminance
measurement.
ISRn
Sets measurement range to the manual, and starts integral illuminance
measurement.
n: 1 6
ISTR0
Sets measurement range to the auto, and obtains the current status data
(illuminance, integral illuminance, integration time, and measurement range)
ISTRn
Sets measurement range to the manual, and obtains the current status data
(illuminance, integral illuminance, integration time and measurement range)
n: 1 6
IEND
Stops integral illuminance measurement.
CA
Starts a zero adjustment.
WHO
Reads product name (IM-600 is read out for IM-600/IM-600M).
VER
Reads software version.
SRL
Reads product serial number.
SCCF_####
Writes correction factor (C.C.F.) into EEPROM of the IM-600.
Enters correction factor with decimal notation #### or #.### or exponent
notation #.###E±##
Range to enter: 0.001 1000
RCCF
Reads correction factor (C.C.F.) #.###E±##
The “_” mark means a space. #### is a numerical value.
When PC send a communication command to the IM-600/IM-600M, the IM-600/IM-600M returns
OK as a receipt acknowledgment. When the IM-600/IM-600M receive improper command, the IM-
600/IM-600M returns NG.
No command can be accepted during zero adjustment. Send commands after
completing the zero adjustment.