Loading...
41
4.1.5 CA COMMAND
When the IM-600/IM-600M receive this command, the IM-600/IM-600M starts zero adjustment.
When IM-600/IM-600M receives this command during integral illuminance measurement, entering
reference value, or entering C.C.F, the IM-600/IM-600M returns FAILED.
PC Command IM-600/IM-600M
CA+(CrLf)
OK+(CrLf)
Finish zero adjustment
SUCCESS+(CrLf)
or FAILED+(CrLf)
4.1.6 WHO/VER/SRL COMMAND
When the IM-600/IM-600M receive this command, The IM-600/IM-600M returns the name of this
instrument, program version, and production serial number.
PC Command IM-600/IM-600M
WHO/VER/SEL+(CrLf)
OK+(CrLf)
In instrument name/version/production serial number
Data +(CrLf)
Loading...