Loading...
42
4.1.7 SCCF COMMAND
When the IM-600/IM-600M receives this command, the IM-600/IM-600M writes C.C.F. into EEPROM
of the IM-600/IM-600M.
Valid values range from 0.001 to 1000.
When a value is over or under 0.001-1,000, the IM-600/IM-600M returns the error code ERR_07.
If the IM-600/IM-600M fails to store the C.C.F. data in the EEPROM, the IM-600/IM-600S returns
code E8.
PC Command IM-600/IM-600M
SCCF_ #### +(CrLf)
OK+(CrLf)
Complete storing C.C.F.
SUCCESS+(CrLf)
or ERR_07+(CrLf)
or ERR_08+(CrLf)
4.1.8 RCCF COMMAND
When the IM-600/IM-600M receive this command, IM-600/IM-600M returns C.C.F. data.
PC Command IM-600/IM-600M
RCCF +(CrLf)
OK+(CrLf)
C.C.F. Data +(CrLf)