5.2 COMMUNICATION ERROR CODE
When an error occurs on the instrument, which communicate with PC, following error messages is
sent to your PC.
Error code
Explanation
Action
ERR01_Rx
Zero adjustment does not complete
adequately.
Check that the detector is
connected to the display unit
correctly. If no problems are
found, contact your dealer or
TOPCON TECHNOHOUSE.
ERR02_Rx
The measured value exceeds measurable
range.
Check the measurement range.
ERR04_Rx
The value reflected by correction factor
exceeds the display range 999900 lx.
Check the measurement range
and use upper range.
ERR05_Rx
Measured vale of integral illuminance
exceeds 1,000,000,000 lxh.
Do not exceed the limit of integral
illuminance.
ERR06_Rx
Integration time exceeds 9999 hours.
Do not exceed the limit of integral
time.
ERR07
The values of reference data, distance data,
or correction factor exceed the limit of entry.
Check the entry value.
ERR08
The instrument fails to write or read data.
Contact your dealer or TOPCON
TECHNOHOUSE.
ERR10
The response switch is pushed 3 times or
more during a zero adjustment.
Don’t touch the response switch
during a zero adjustment.
ERR11_Rx
STRn or ISTRn command is sent to the IM-
600 while HOLD switch is ON and
measurement range is being changed.
Check that the hold switch is
turned OFF during
communication.
ERR12
Commands are sent during a zero
adjustment.
Send commands after completing
the zero adjustment.
NG
A command is not proper.
Check that the entered command
is correct.
FAILD
Command for start or stop of integration time,
or command for a zero adjustment is failed.
Start/stop of integration time or
zero adjustment cannot be set
during integral illuminance
measurement. Try it again after
completing integration illuminance
measurement.