5. Specifications
Pressure Module Dependent Operates with selected PX971
Ranges: Transducers
T/C Ranges: J, K, T, E, R, S, N, B, L, U including – 10 to 70mV
RTD Ranges: Pt 385 (100, 200, 500, 1000 ohms)
Pt 392, JIS, Ni120, CU10, YSI400
Ohms Ranges: 0 to 400.00 and 400.0 to 4000.0
mA Range: 0 to 24.000 mA
Voltage: (Low) Input 0-20.000 V, Output 0-10.000 V
(High) Input only 0-250 VDC or VAC
Frequency: 0-1000.0 CPM
0-1000.0 Hz
0-10.00 KHz
Continuity: 0 to 400.0 ohms, audible tone when under 10 ohms
Accuracy: T/C Type J: ±.3°C ±10 mV (±0.6°C total error)
RTD PT100: ±.1°C ±.075 W (±0.3°C total error)
Ohms (400): ±0.1 W
Ohms (3200): ±1.0 W
mA Compliance: Up to 1000 W load
Voltage Drive
Capability: 1 mA
RTD IEX-range: 0.01 mA to 3 mA
RTD Frequency
Response: 10 mS – works with all pulsed XMTR’s
Oper. Temp: -10°C to 50°C
Storage Temp: -40°C to 60°C
Loop Power: 4 AA alkaline batteries, 25 hours typical operating time
Size: 7.5" x 4" x 1.5"
Range & Accuracy
Range Min Max Accuracy
(% of reading ± counts)
V Read (low) 0.000 20.000 0.015% ± 2
VDC Read (high) 0.0 250.0 0.05% ± 2
VAC Read (high) 0.0 250.0 0.5% ± 2
V Source 0.000 10.000 0.015% ± 2
mV Read 0.00 90.00 0.015% ±2
mV Source 0.00 100.00 0.015% ±2
mA Read 0.000 24.000 0.015% ± 2
mA Source 0.000 24.000 0.015% ± 2
CPM Source 1.0 600.0 ± 0.05%
Hz Source 1 1000 ± 0.05%
KHz Source 1.00 10.00 ± 0.25%
CPM Read 0.0 600.0 0.05% ± 1
Hz Read 0 1000 0.05% ± 1
KHz Read 0.00 10.00 0.05% ± 1
Note: The frequency output is a 5