11
1GENERAL OVERVIEW
1.1 Description
PLANAR 804/1 Network Analyzer is designed for use in the process of development,
adjustment and testing of various electronic devices in industrial and laboratory
facilities, including operation as a component of an automated measurement system.
PLANAR 804/1 is designed for operation with external PC, which is not supplied with
the Analyzer.
1.2 Specifications
1.2.1 Basic Specifications
Table 1.1 Basic Specifications
1 2
Frequency range 300 kHz to 8.0 GHz
CW frequency accuracy ±5´10–6
Harmonic distortion –25 dB
Non harmonic spurious –30 dB
Output power level
-frequency range 300 kHz to 6.0 GHz
- frequency range 6.0 GHz to 8.0 GHz
–60 dBm to +10 dBm
–60 dBm to +5 dBm
Output power level accuracy ±1.0 dB
Magnitude transmission measurement accuracy1, if ½S11½
and ½S22½ of the DUT are less than –32 dB, and ½S21½
and ½S12½ values are as follows:
+5 dB to +15 dB
–50 dB to +5 dB
–70 dB to –50 dB
–90 dB to –70 dB
0.2 dB
0.1 dB
0.2 dB
1.0 dB
1The specifications of the Analyzer apply over the temperature range of 23 °C ±5 °C (unless
otherwise specified) after 40 minutes of warming-up, with less than 1 °C deviation from the full two-port
calibration temperature, at output power of –5 dBm, and IF bandwidth 1 Hz.