5 CALIBRATION AND CALIBRATION KIT
114
5.2 Calibration Procedures
5.2.1 Calibration Kit Selection
The calibration kit employed during a calibration should be selected according to the
following procedure. If it is not specified in the list of the predefined calibration kits,
you should add it. The procedure of adding and editing of the calibration kits is
described in the section 5.3.
To open the list of the calibration kits (See figure 5.6), use
the following softkeys:
Calibration > Cal Kit
Figure 5.6 The list of calibration kits
Highlight the required line in the list of the calibration kits
and use the following softkey:
Select
Or click on the checkbox in the row "Select" by the mouse.
Note Make sure that the selected calibration kit is checkmarked.