5 CALIBRATION AND CALIBRATION KIT
119
5.2.4 Full One-Port Calibration
Full one-port calibration is used for reflection coefficient measurements (S11 or S22). The
three calibration standards (SHORT, OPEN, LOAD) are measured (See figure 5.9) in
the process of this calibration.
Figure 5.9 Full one-port calibration
Before starting calibration perform the following settings: select active channel, set the
parameters of the channel (frequency range, IF bandwidth, etc), and select the
calibration kit.
To open full one-port calibration submenu, use the
following softkeys:
Calibration > Calibrate > Full 1-Port Cal
Select the test port to be calibrated using Select Port.
Clicking this softkey you can switch between the test ports
(measured parameters).
Connect SHORT, OPEN and LOAD standards to the
selected test port in any consequence as shown in figure
5.9. Perform measurements clicking the softkey
corresponding to the connected standard.
The instrument status bar will indicate Calibration in
progress... when the measurement is in progress. On
completion of the measurement a check mark will appear
in the left part of the softkey.
Port
SHORT
LOAD
OPEN