5 CALIBRATION AND CALIBRATION KIT
121
5.2.5 One-Path Two-Port Calibration
One-path two-port calibration is used for measurements of the DUT parameters in one
direction, e.g. S11 and S
21. This method involves connection of the three calibration
standards to the source port, and connection of a THRU standard between the calibrated
source port and the other receiver port (See figure 5.10). You can also perform isolation
calibration by measuring two LOAD standards.
Figure 5.10 One-path two-port calibration
Before starting calibration perform the following settings: select active channel, set the
parameters of the channel (frequency range, IF bandwidth, etc), and select the
calibration kit.
To open one-path two-port calibration submenu, use the
following softkeys:
Calibration > Calibrate > One Path 2-Port Cal
Port
1
Port
2
LOADs
(optional)
Port
1
2
Port
SHORT
LOAD
OPEN